title: Depyrogenation Ovens & Tunnels EQ&C | Mesa Labs
description: Mesa’s Compliance and Validation Services can ensure that your Depyrogenation Oven/Tunnel meets your depyrogenation/sterilization requirements.
image: https://lynton-sb3-mesalabs-20969579.hs-sites.com/hubfs/mesalabs%20website/solutions/solutions/continuous%20and%20process%20monitoring/Validation%20and%20Calibration%20Services/Validation%20and%20Calibration%20Services%20Landing/PQServices%26Documentation.png
---

# Depyrogenation Ovens & Tunnels EQ&C

1. [Home](https://lynton-sb3-mesalabs-20969579.hs-sites.com)
2. caret-right icon
   
   [Our Solutions](https://lynton-sb3-mesalabs-20969579.hs-sites.com/solutions)
3. caret-right icon
   
   <https://lynton-sb3-mesalabs-20969579.hs-sites.com/>
4. caret-right icon
   
   [Services](https://lynton-sb3-mesalabs-20969579.hs-sites.com/services/)
5. caret-right icon
   
   Depyrogenation Ovens & Tunnels EQ&C

![PQ Services & Documentation](https://lynton-sb3-mesalabs-20969579.hs-sites.com/hs-fs/hubfs/website/solutions/solutions/continuous%20and%20process%20monitoring/Validation%20and%20Calibration%20Services/Validation%20and%20Calibration%20Services%20Landing/PQServices%26Documentation.png?width=346&height=300&name=PQServices%26Documentation.png)

When it comes to completing Depyrogenation Oven/Tunnel Installation, Operational, or Performance Qualifications, the protocols used need to verify that the Depyrogenation Oven/Tunnel has been installed correctly (IQ), demonstrate that the Depyrogenation Oven/Tunnel is operating properly (OQ), and show that the Depyrogenation Oven/Tunnel sterilization cycles are able to reproducibly meet your requirements (PQ).

#### Mesa’s Compliance and Validation Services Depyrogenation Oven / Tunnel Qualification protocols include:

- Document the Depyrogenation Oven/Tunnel Information (Make, Model, Serial Number, etc.). (IQ)
- Identification and Verification of all Critical Instruments used by the Depyrogenation Oven/Tunnel. (IQ)
- Verification that the Depyrogenation Oven/Tunnel Controls and Alarms operate as required by the customer and per the manufacturer’s specifications. (OQ)
- Verification that Standard Operating Procedures are available for the operation, maintenance, and/or cleaning of the Depyrogenation Oven/Tunnel.
- Verification that the Depyrogenation Oven/Tunnel is able to meet and maintain the required temperature during the depyrogenation/sterilization cycle. (OQ)
- Verification that the Depyrogenation Oven/Tunnel depyrogenation/sterilization cycles can reproducibly meet the requirements for time, temperature, pressure, sterility assurance using endotoxin or biological indicator reduction. (PQ)
- Development of different Depyrogenation cycles to meet your needs & requirements for time, temperature, and endotoxin or biological indicator reduction.

By using Mesa’s Compliance and Validation Services to complete your Depyrogenation Oven / Tunnel qualifications, the completed IQ, OQ, or PQ protocols will provide a level of assurance that the Depyrogenation Oven/Tunnel has been installed properly, is operating correctly and is able to reproducibly meet your (the customer’s) depyrogenation/sterilization requirements.